Air Shields Transport Incubator Gurney Stretcher for Sale

Air Shields Transport Incubator Gurney Stretcher for Sale

We have this Air Shield incubator in stock along with photo-therapy lights, infant warmers, infant care centers and other neo-natal monitors and hospital equipment. "Hospital Direct Medical Equipment" 4901 Morena Blvd. #506 San Diego Ca. 92117 getinfo@san.rr.com www.1usedmedicalequipment,.com




FEATURES

  • Intended for transport of high risk, premature, low birth-weight, or critically ill newborns.
  • It provides a means to control air temperature, oxygen concentrations, and relative humidity.
  • A double wall hood permits full visibility and provides an effective thermal barrier from the environment.

 

SPECIFICATIONS

Dimensions (with stand):
  • Length: 40.25 in. (102 cm)
  • Width: 22.25 in. (56.5 cm)
  • Height: 46.75 in. (118.7 cm) maximum, 34.75 in. (88.3 cm) minimum
Power Requirements:
  • AC: 110.120 VAC 50/60/400 Hz, sine or square wave 170/200 Watts and 225/270 Watts with batteries charging.
  • DC: 12 VDC, 115 Watts
Internal Battery:
  • Type: gel type, sealed and rechargeable.
  • Voltage: 12 VDC nominal
  • Quantity: 2
  • Charging Time: two batteries take 20 minutes to recharge from fully discharged.
  • Life expectancy: minimum 200 full charge/discharge cycles.
Chassis Leakage Current: 100µA or less

Alarms:
  • High Temp: Actuates at displayed incubator temperature of 39º ± 0.5 ºC.
  • Sensor: Actuates if High Temp alarm sensor is electrically open or shorted.
  • Power Failure: Actuates if main AC power fails and no DC back-up power is present.
  • Heater Temp: Actuates if heater temperature exceeds 77 ºC nominal.
  • Low DC: Actuates if incubator DC voltage has fallen below 10.5 VDC nominal.
Temperature:
  • Set Point Range: 21.5 ºC ± 1.0 ºC to 38 ºC
  • Rise Time: 30 minutes
  • Variability: 0.2 ºC
  • Overshoot: 1.5 ºC maximum
  • Uniformity: 1 ºC
  • Correlation of Displayed Incubator to Actual Incubator Temperature at Temperature Equilibrium: 1 ºC
  • Correlation of Displayed Incubator Temperature to Set Point at Temperature Equilibrium: ± 0.5 ºC
  • Observation Light Output: 10-foot candles at mattress level
Hospital Direct Medical Equipment 4901 Morena Blvd. #506 San Diego, CA 92117 Phone: (858) 263-4894 Email: hospitaldirectmedical@gmail.com

Comments

Medical Equipment Blog Posts

Operating Room Equipment Company

Operating Room and Surgical Equipment Company

Stryker 1015 SM204 Surgery Stretcher Overview

Stryker Eye Surgery Stretchers Used Refurbished

Exam Tables Exam Room Medical Equipment San Diego CA

Used Ultrasound Tables

Stryker 1115 Prime Series Stretcher Model

Ritter 223 Exam Table

Midmark Ritter 104 Exam Table

Midmark 622 Exam Table